DRK108A Paper Tearness Tester

Short Description:

DRK108A Paper Tearness Tester is a special instrument for the determination of tear strength. This instrument is mainly used for the determination of paper tearing, and it can also be used for the determination of tearing of lower-strength cardboard.


Product Detail

Product Tags

DRK108A Paper Tearness Tester is a special instrument for the determination of tear strength. This instrument is mainly used for the determination of paper tearing, and it can also be used for the determination of tearing of lower-strength cardboard. It is used for papermaking, packaging, scientific research and product quality. Ideal laboratory equipment for supervision and inspection industries and departments.

Features
Compact structure, convenient operation, beautiful appearance, convenient maintenance; multi-function, flexible configuration,
The measurement results are directly obtained, and the instrument is precisely manufactured and has high measurement accuracy.

Applications
The instrument is mainly used for paper tear measurement. Changing the configuration of the instrument can be widely applied to the measurement of other materials, such as plastic, chemical fiber, metal wire, and metal foil.

Technical Standard
GB/T450-2002 “Taking samples of paper and cardboard (eqv IS0 186: 1994)”
GB/T10739-2002 “Standard Atmospheric Conditions for the Processing and Testing of Paper, Paperboard and Pulp Samples (eqv IS0 187: 1990)”
ISO1974 “Paper—Determination of Tearing Degree (Elimendorf Method)”
GB455.1 “Determination of paper tearing degree”

Product parameter

Project Parameter
Standard pendulum measurement range (10—1000)mN division value 10mN
Light pendulum (10~1000)mN, division value 5mN (optional)
Lightest pendulum (10~200)mN, division value 2mN (optional)
Indication error ±1% within the range of 20%-80% of the upper limit of measurement, ±0.5% FS outside the range.
Repeatability error 20% of the upper limit of measurement—80% within the range <1%, outside the range <0.5% FS
Tear arm (104±1)mm
Initial angle of tear 27.5°±0.5°
Tear distance (43±0.5)mm
Paper clip surface size (25×15) mm
Distance between paper clamps (2.8±0.3)mm
Sample size Should be (63±0.5)mm×(50±2)mm
Working conditions Celsius temperature: 23, relative humidity 50%+/-5
Dimensions 420×300×465mm
Quality 25㎏.

Product Configuration
One host, one manual.


  • Previous:
  • Next:

  • Write your message here and send it to us

    Products categories